Syringe Pump

Syringe Pump - Crest

Product Details:
  • Dimension : 130 mm (W) x 183mm (L) x 115mm (H)
  • Power Source : 110/220 VCA (+/-20%), 50/60 Hz, 12 VA
  • Operating Temp : 05 deg - 40 deg C
  • Weight: 1.8 Kgs
  • Display Size & Type : 135mm x 21mm EBTN LCD Display
  • Relative Humidity: 15 to 90% (no condensation)
  • Atmospheric Pressure: 70 to 106 Kpa
  • Battery: 4.8 Rechargeable NiMh Capacity minimum 4hrs @ 5ml/hr Recharging time 16 hrs
  • Syringe Size: 10 ml-10 ml, 10 ml-60 ml, 60 ml-100 ml, 10 ml-10 ml, 10 ml-60 ml, 60 ml-100 ml
  • Number of Syringes: 1.0
  • Minimum Order Quantity: 5 Unit

Physical Specification:
  • Product Type: Front Loading Syringe Infusion pump
  • Pumping Mechanism: Lead Screw Mechanism
  • Parameters Diplayed: Syringe size, Syringe Brand, Set Flow Rate Infused Volume, Occlusion level, Battery life Alarms & Indicators
  • Keypad: Capacitive touch screen keypad
  • Fixation: Allows Horizontal and Vertical Mounting
  • Stackability: For interlocking of two or more units

Basic Infusion Controls:
  • Syringe Capacity: 5,10,20,30,50/60 and 100 CC Auto Detection
  • Syringe Brands: Dispovan, BBraun Perfusor, BD Precise, Platipak Romson & Sisco (Provision add 12 more brands)
  • Flow Rate: 0.1 to 1200 ml/hr (increment in step of 0.1ml/hr Facility for Online titration
  • Purge: Upto 1200 ml/hr
  • Bolus Rate: Upto 1200 ml/hr
  • Bolus Volume: Settable (pre-set)
  • KVO/KOR: 1% of the set rate or minimum of 0.1 ml/hr
  • Mechanical Accuracy: +/- 1%
  • Overall Accuracy: +/- 2% (Measured Time > 1Hr, min 0.2 ml excluding syringe tolerance)
  • (Measured as per IEC 60601-2-24)

Pressure Management:
  • Pressure Limits: Low 0.4 kg/cm2 (300 mmHg), Mid 0.8 kg/cm2 (600 mmHg), High 1.2 kg/cm2 (900 mmHg)
  • Tolerance: +/- 20%
  • Anti-bolus System: Reduces significantly bolus after occusion release (measured as per IEC 60601-2-24)

Pre Alarms/Alarms/Security Features:
  • Alarms: All the alarms are expressed by means of flashing message/symbol on the display and sound beep
  • Syringe Detection: Syringe Dislocated, Drive Disengaged, Patient Line Disconnection, Plunger Disengaged
  • Device Control: No Mains, Battery pre-alarm & Alarm, Idle Technical Malfunction
  • Infusion Control: Occlusion, Infusion End pre-alarm & Alarm KVO/KOR, Syringe Empty

Standard Accessories:
  • Unit with pole clamp, power cord & operating manual

Optional Feature:
  • Power Source: External Power 12 VDC converter (For Ambulatory supply)
  • Battery Life: Upgradable to 5 hr / 10 hr @ 5 ml/hr
  • Stackability: Multiple inputs AC cord for up to 8 pumps
  • Communication: RS232 port
  • Other: Nurse call attachment

Quality Standard:
  • Electrical Safety: Class II, Type CF

Additional Information:
  • Item Code: CREST

Syringe Pump - Zeta

Product Details:
  • Dimension: 130mm (W) x 183mm (L) x 115mm (H)
  • Weight: 1.8 Kgs
  • Display Size & Type: 66mm x 33mmm Graphic LCD Monochrome
  • Power Source: 110/220 VCA (+/- 20%), 50/60 Hz, 12 VA
  • Battery: 4.8 Rechargeable NiMh Capacity minimum 4 hrs @ 5 ml/hrRechargingg time 16 hrs
  • Operating Temp.: 05 to 40 deg C
  • Atmospheric Pressure: 70 to 106 Kpa
  • Relative Humidity: 15 o 90% (no condensation)
  • Application: Drug Delivery
  • Syringe Size: 5 ml - 100 ml, 10 ml-10 ml, 10 ml-60 ml, 60 ml-100 ml
  • Number of Syringes: 1.0
  • Minimum Order Quantity: 5 Unit

Product Specification:
  • Product Type: Front Loading Syringe Infusion Pump
  • Pumping Mechanism: Lead Screw Mechanism
  • Parameters Displayed: Occlusion Level, Syringe Size, Syringe Brand, Drug Name, Set Flow Rate, Infused Volume, Battery Life, Alarms & Indicators
  • Keypad: Capacitive Touch Screen Keypad
  • Fixation: Allows Horizontal & Vertical Mounting
  • Stackability: For Interlocking of two or more units
  • Battery Status: Display in Bars OR hrs, and min.
  • Power Saver: Auto adjust display brightness to save battery

Basic Infusion Controls:
  • Syringe Capacity: 5,10,20,30,50/60 and 100 CC Auto Detection
  • Syringe Brands: Dispovan, BBRaun Perfusor, BD Precise, Plastipak Romson & Sisco (Provision to add 13 more brands)
  • Flow Rate: 0.1 to 2000 ml/hr (increment in step of 0.1 ml/hr) Facility for online titration
  • Purge: Upto 2000 ml/hr
  • Priming: 3Modes: Mandatory/ Non Mandatory/ Advised Rate: Upto 2000 ml/hr
  • Bolus Type: Auto & Manual
  • Bolus Rate: Upto 2000 ml/hr
  • Bolus Volume: 0.1 ml to max capacity of syringe (0.1ml incr.)
  • Bolus Indicator: Beep after infusion of every 0.5/1 ml bolus volume
  • Infusion Modes: 1: Rate ml/hr (0.1ml /hr incr., 2. Volume Limit 0.1-999.9 ml (01 ml incr.), 3. Volume /Time 0.1-999.9 ml; 00h01-100h00, 4. Dose Rate, # Dose Unit: mcg/min, mcg/hr, mcg/kg/min mcg/kg/hr mg/min, mg/hr mg/kr/min, mg/kg/hr, # Dilution: X mg/ml or X mg / Y ml, 5. Upgradable to PCA
  • Standby: Programmable form 1 min. to 24 hrs.
  • Drug Labelling: 64 drug name with provision for user add/delete
  • KVO/KOR: Settable from 0.1 to 5 ml/hr Provision to keep it off.
  • Mechanical Accuracy: +/- 1%
  • Overall Accuracy: +/- 2% (Measured time, 1 Hr, min 0.2 ml

Pressure Management:
  • Pressure Modes: Variable / 3 Pre-set Levels /10 Pre-set Levels From 100 to 900 mmHg, 50 mmHg increment
  • Pressure Limits: Low 0.4 kg/cm2 (300 mmHg)
  • Mid 0.8 kg/cm2 (600 mmHg)
  • High 1.2 kg/cm2 (900 mmHg)
  • Tolerance: +/- 20%
  • DPS: Warns Pressure Variation
  • Pressure Monitoring: Graphic Representation of the pressure limits
  • Anti-bolus System: Reduces significantly bolus after occlusion release

Pre Alarms/ Alarms/ Security Features:
  • Alarms: All the alarms are expressed by means of flashing message/symbol on the display and sound beep
  • Syringe detection: Syringe Dislocated, Drive Disengaged, Patient Line Disconnection, Plunger Disengaged
  • Device Control: No Mains, Battery Pre-alarm & Alarm, Idle, Technical Malfunction, Unconfirmed Programming Preventive Maintenance
  • Infusion control: Occlusion Pre-alarm & Alarm, Infusion End Pre-alarm & Alarm, KVO/KOR, Syringe Empty
  • Keypad Locking: Restricts Accidental change of the infusion parameters

Standard Accessories:
  • Unit with pole clamp, power cord & operating manual

Optional / Upgradable Feature:
  • Power Source: External Power 12 VDC converter (For ambulatory power supply)
  • Battery Life: Upgradable to 5/10 hr @ ml/hr
  • Stackability: Multiple input AC cord to power upto 8 pumps
  • Communication Port: Rs232 port
  • Data Log Event: 500 data log events in real-time Expandable upto 1500 nos.
  • Other: Nurse call attachment

Syringe Pump - Zion

Product Details:
  • Syringe Detection: Automatic
  • Power Source: 110/220 VAC (+/-10%), Freq: 50/60Hz, 12 VA, External power 12 VDC by a converter.
  • Dimensions: 135 (H) x 272 (W) x 150 (D) mm
  • Weight: 2.3 Kgs./ Housing: ABS Plastic
  • Communication Port: RS 232C
  • Syringe Loading: Front side
  • Display: Graphic LCD monochrome, size 66 mm x 33 mm ( 128 x 64 pixels)
  • Application: Drug Delivery
  • Syringe Size: 5 ml - 100 ml, 10 ml-10 ml, 10 ml-60 ml, 60 ml-100 ml
  • Number of Syringes: 1.0
  • Minimum Order Quantity: 5 Number

Infusion:
  • Flow Rate Range: 0.1 to 1200 ml/hr in normal mode, depending on syringe capacity (0.1 ml/hr increment)
  • Syringe Capacity: 05, 10, 20, 30/35, 50/60 ml
  • Type Of Syringe: Dispovan, BD Precise, Plastipak, Bbraun Perfusor, Romoson ( Capacity to store 20 brands)
  • Priming: Priming Rate: Up to 1200 ml/h depending on syringe size
  • Bolus: Bolus rate: 1 to 1200 ml/hr (1 ml/hr Increment), Bolus volume: 0.1 ml to the max capacity of syringe ( 0.1 ml increment)
  • Infusion Modes: 1] ml/h (0.1 ml/h increments), 2] Volume limit: 0.1-999.9 ml (0.1 ml increments), 3] Volume / time: 0.1-999.9 ml; 00h01- 100h00., 4] Dose rate., # Dose units: mcg, mcg/min, mcg/hr, mcg/kg/min, mcg/kg/hr,mg, mg/min, mg/hr, mg/kg/min, mg/kg/hr, # Dilution setting: X mg/ml or X mg / Y ml.
  • Drug Labelling: Display of the available drug name from the drug library. The user can add or delete drug names up to 64 nos.
  • Accuracy: Mechanical accuracy: ±1%, Overall Accuracy: ±2%,(excluding syringe tolerances)
  • End Of Infusion: (V/T & VL) KVO 1 and KVO 2: from 0.1 to 5 ml/h adjustable and storable, continuous infusion or stop.

Pressure Management:
  • Pressure Modes: Low:- 0.4 kg/cm2 (300 mmHg), Mid:-0.4 kg/cm2 (600 mmHg), High:-0.4 kg/cm2 (900 mmHg), tolerance +/-20%
  • Anti-Bolus System: Reduces significantly bolus after occlusion release.

Pre-Alarms/ Alarms/ Security:
  • Pump Status: Red for AC power status and Green for infusion in progress. Idle/ Nonoperative time over. Visibility up to 5 meters. All alarms are expressed by means of flashing messages on LCD, graphics and sound beep.
  • Syringe Position Control: Syringe barrel knob check, Plunger head detection, Anti-siphon system check.
  • Device control: Disengaged driving mechanism alarm, Low battery pre-alarm, Discharged battery alarm, Unconfirmed programming, Technical malfunction alarm, Drive system advance check, Watchdog check.
  • Infusion Control: Occlusion pressure alarm, End of infusion pre-alarm, End of infusion alarm, Syringe empty, KVO, Volume limit pre-alarm, Volume limit alarm, keypad locking.

Technical Specifications:
  • Battery: 8.4V Rechargeable Ni-Mh, Capacity minimum:10 hrs @ 5 ml/hr, Battery replacement: separate compartment – no need for electrical security testing.
  • Operating Conditions: Temp: 05° to 40°C/ Humidity: 15 to 90 % (no condensation)/ Atmospheric Pressure: 70 to 106 kPa
  • Stackability: Electrical: AC Power cord with multiple inputs to connect max 8 devices on single power socket.
  • Standard Accessories: Unit with IV Pole Mounting Device, Operating Manual

Syringe Pump - Zion Plus

Product Details:
  • Display: Graphic LCD monochrome, size 66 mm x 33 mm ( 128 x 64 pixels)
  • Power Source: 110/220 VAC (+/-10%), 50/60Hz, 12 VA, external power 12 VDC by a converter.
  • Dimensions: 135 (H) x 272 (W) x 150 (D) mm / Weight: 2.3 Kgs./ Housing: ABS Plastic.
  • Communication Port: RS 232C
  • Stackability (Optional): Up to 3 devices self - stackable on a pole
  • Standard Accessories: Unit with IV Pole Mounting Device, Operating Manual
  • Application: Drug Delivery
  • Number of Syringes: 1.0
  • Minimum Order Quantity: 5 Number

Infusion:
  • Flow Rate Range: Up to 2000 ml/hr in normal mode, depending on syringe capacity (0.1 ml/hr increment)
  • Syringe Capacity: 05, 10, 20, 30/35, 50/60 ml
  • Type Of Syringe: Dispovan, BD Precise, Plastipak, BBraun Perfusor, Romoson ( Capacity to store 20 brands)
  • Priming: 3 modes: mandatory, not mandatory, or advised / Rate: Up to 2000 ml/h depending on syringe size
  • Bolus: Bolus rate 1 to 2000 ml/hr (1 ml/hr Increment) . Bolus volume: 0.1 ml to max capacity of syringe (0.1 ml increment)
  • Infusion Modes: 1] ml/h (0.1 ml/h increments), 2] Volume limit: 0.1-999.9 ml (0.1 ml increments), 3] Volume / time: 0.1-999.9 ml; 00h01- 100h00., 4] Dose rate. # Dose units: mcg, mcg/min, mcg/hr, mcg/kg/min, mcg/kg/hr,mg, mg/min, mg/hr, mg/kg/min, mg/kg/hr, U, kU, # Dilution setting: X mg/ml or X mg / Y ml.
  • Drug Labelling: Display of the available drug name from the drug library. The user can add or delete drug names up to 64 nos.
  • Drug Library (Optional): Inbuilt ‘PRUDENT’ Drug library editor for setting normal infusion, bolus, and loading dose parameters. Adjustment of the dilution, default infusion modes, flow rate and bolus limits in ml/h.
  • Accuracy: Mechanical accuracy: ±1%, Overall Accuracy: ±2%,(excluding syringe tolerances)
  • End Of Infusion: (V/T & VL) KVO: from 0.1 to 5 ml/h adjustable and storable, continuous infusion or stop.
  • Pause: Programmable from 1 minute to 24 hours, increments from minute to minute.
  • Data Log Event: 500 datalog events in real time expandable up to 1500 nos.

Pressure Management:
  • Pressure Modes: 3 modes available: variable or 3 pre-set levels or 10 pre-set levels - from 100 to 900 mmHg, 50 mmHg increment Low:- 0.4 kg/cm2 (300 mmHg), Mid:-0.4 kg/cm2 (600 mmHg), High:-0.4 kg/cm2 (900 mmHg), tolerance +/-20%
  • DPS: The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
  • Pressure Monitoring: Graphic representation of the pressure limit
  • Anti-Bolus System: Reduces significantly bolus after occlusion release.

Pre-Alarms/ Alarms/ Security:
  • Pump Status: Red for AC power status and Yellow for infusion in progress. Idle/ Nonoperative time over. Visibility up to 5 meters. All alarms are expressed by means of flashing messages on LCD, graphics and sound beep.
  • Syringe Position Control: Syringe barrel knob check, Plunger head detection, Anti-siphon system check, Flange detection.
  • Device Control: Disengaged driving mechanism alarm, Low battery pre-alarm, Discharged battery alarm, battery capacity, display in hours and minutes, unconfirmed programming, Technical malfunction alarm, Drive system advance check, Watchdog check, Communication
  • connection failure.
  • Infusion Control: Occlusion pressure pre-alarm, Occlusion pressure alarm, End of infusion pre-alarm, End of infusion alarm, Syringe empty, KVO, Volume limit pre-alarm, Volume limit alarm, Keypad locking, Hard and soft flow rate limits.
  • Maintenance: Preventive maintenance warning.

Technical Specifications:
  • Battery: 8.4V Rechargeable Ni-Mh, Capacity minimum:10 hrs @ 5 ml/hr,Battery replacement: separate compartment – no need for electrical security testing.
  • Operating Conditions: Temp: 05° to 40°C/ Humidity: 15 to 90 % (no condensation)/ Atmospheric Pressure: 70 to 106 kPa
Looking for Syringe Pump ?
Share Us:
© Plenum Tech Private Limited
Desktop Site Back to top